Technical Specifications: Temperature Calibrator
|
Analogue Input Ranges |
Impedance |
Accuracy |
Resolution |
0 to 24 mA |
Input Impedance ± 17 Ω, 250mA Fused |
0.02 % of FS |
1 µA |
-10 to 100 mV |
Input Impedance > 1 MΩ |
0.01 % of FS |
1 µA |
Tc Types K, J, T, E, N, U and L |
Input Impedance > 1 MΩ |
0.1 % of FS* |
0.1 °C |
Tc Types B, R and S |
Input Impedance > 1 MΩ |
1 % of FS* |
0.1 °C |
0 - 400.00 Ω |
|
0.05 % of FS |
0.01 Ω |
0 - 2200.0 Ω |
|
0.05 % of FS |
0.1 Ω |
RTD types Pt50, Pt100, Pt500, Pt1000, Ni100, Ni120 |
|
0.1 % of FS |
0.01 °C |
|
Measurement of Ohms and RTD's in 2, 3 or 4-wire configurations |
Continuity with a less than 100 Ω trigger and 1mA excitation, with visual and audible confirmation |
|
Analogue Output Ranges |
Load |
Accuracy |
Resolution |
0 to 24 mA |
Output load max 500 Ω |
0.02 % of FS |
1 µA |
-10 to 100 mV |
Output load min 100 Ω |
0.01 % of FS |
1 µV |
Thermocouple Types K, J, T, E, N, U and L |
0.1 % of FS* |
0.1 °C |
Thermocouple Types B, R and S |
1 % of FS* |
0.1 °C |
1 to 400.00 Ω |
10 to 400 Ω |
0.1 % of FS |
0.01 Ω |
1 to 2200.0 Ω |
10 to 2200 Ω |
0.05 % of FS |
0.1 Ω |
RTD types Pt50, Pt100, Pt500, Pt1000, Ni100, Ni120 |
0.1 % of FS |
0.01 °C |
|
* Excluding cold junction, using included Thermocouple Extender Module with CJC set to External |