Technical Information
|
Superdew 3
|
Sensor |
SHAW Ultra High Capacitance Aluminium Oxide Sensor |
Ranges |
P (Purple): -100 °C to 0 °C dewpoint 0-6,000 ppm(V) S (Silver): -100 °C to -20 °C dewpoint 0-1,000 ppm(V) G (Grey): -80 °C to 0 °C dewpoint 0-6,000 ppm(V) R (Red): -80 °C to -20 °C dewpoint 0-1,000 ppm(V) B (Blue): -80 °C to +20 °C dewpoint 0-23,000 ppm(V) |
Approvals |
CE and UKCA |
AutoCal |
Field calibration/span check facility. Activated and operated via potentiometer on the front of the instrument. |
Factory Calibrations |
Supplied with Certificate of Calibration traceable to the National Physical Laboratory (NPL) |
Accuracy |
±2 °C (±3.6 °F) dewpoint NPL traceable -90 °C to +20 °C (-130 °F to +68 °F) |
Reasolution |
0.1 °C, 0.1 °F dewpoint or 0.1 ppm(v) |
Repeatability |
Better than ±0.3 °C (±0.54 °F) dewpoint |
Operating Temperature |
-20 °C to +60 °C (-4 °F to +140 °F) |
Storage Temperature |
-20 °C to +70 °C (-4 °F to +158 °F) |
Response Times |
Wet to dry: -20 °C to -60 °C - less than 120 secs Dry to wet: -100 °C to -20 °C - less than 20 secs |
Sample Flow Rate |
Flow independent but ideally 2 - 5 litres per minute Max: 25 litres/min |
Calibration Warranty |
12 months from date of dispatch |
Electrical
|
Electromagnetic Compatibility (ECM) |
Immunity: Complies with EN 61326:2006 Emissions: Complies with EN 61326:2006 |
Warm Up Time |
5 seconds |
Fault Conditions |
Short circuit/over range: display reads “Short” and output drives to 24.0 mA |
Output Signal |
Isolated 4 - 20 or 0 - 20 mA Linear |
Operating Voltage |
90 to 250 V AC 50/60 Hz or 24 V DC |
Display |
5 character LED |
Alarms |
Two single pole changeover contacts (NO/C/NC), rated at 10 A at 240 V AC. Alarm trip at set point with 0.1 °C (or equivalent) hysteresis to eliminate relay chatter |
Cable |
Supplied with two metre low loss coaxial cable |
Mechanical
|
Operating Pressure |
From 1kPa (0.01 barA) to maximum 35,000 kPa (350 barA) |
Operating Humidity (External) |
Maximum - 95% RH non-condensing |
Enclosure |
DIN Style, 144 mm wide, 72 mm high by 116 mm deep ABS case |
Mounting |
Panel mount 138.0 x 68.0 mm (DIN 43700) |
Sensor Protection |
316 sintered stainless steel filter - 50 micron |
Probe Material (Wetted Parts) |
316 stainless steel |
Weatherproof Classifications |
Front panel only - IP54 |