Manufacturer:
The GDA 4350 Carbon Dioxide Gas Sensor can detect CO₂ at very low levels, and will allow you to monitor indoor air quality with ease.
Part Number: GDA-4350
Back to Previous Page
Call for pricing

Description

The GDA 4350 Carbon Dioxide Gas Sensor is a 3 wire Non Dispersive Infra red (NDIR) Carbon Dioxide (CO₂) sensor. It gives a 4-20 mA, 0-20 mA, 0-5 V or 0-10 V o/p over the range of the sensor.

Housed in an ABS plastic case with an aperture for the gas path. It uses laser LED to give reliable performance. Calibration for non critical safety applications is every 3 years. The low temperature version (suffix LT) operate from - 25°C to 30°C.


Highlighted Features

  • 24 VDC, 3 Wire Operation
  • 4-20 mA, 0-20 mA, 0-5 V or 0-10 V o/p linear over range of sensor
  • Ultra low cross sensitivity to other gas
  • Temperature and humidity compensated
  • Multiple ranges
  • No moving parts.
  • Factory calibrated
  • Low drift

Designed, manufactured and supported locally in Australia.

Technical Information

Specification NameSpecification Value
Detectable Gases Carbon Dioxide (CO₂)
Detectable Gas Ranges 0-2000 ppm;10000 ppm; 5% vol; 20% vol; 100% vol
Power Requirements 24 VDC
Output Types 4-20 mA, 0-20 mA, 0-5 V or 0-10 V
Power Consumption <30 mA @ 24 VDC
Output Specifications max loop Impedance 800R @ 24 VDC
Wiring 3-wire (0.75mm2) with overall screen
Operating Temperature 5°C to +45°C (LT version to -20°C to +45°C)
Operating Humidity 5% to 95% RH peak non condensing
Operating Pressure 900 - 1100 mbar
Accuracy of Reading +- 2% FSD +-0.5% reading @SATP
Warm Up Time <30 sec
Response Time T90 < 20 sec at SATP
Zero Drift <0.5% FSD for 90days @SATP
Calibration Drift <0.5% FSD for 90 days @SATP
Sensor Technology Non Dispersive Infra-Red (NDIR)
Calibration Requirement Every 3 years with 6 monthly functional tests if sensor operated at SATP
Warranty 12 months from date of dispatch
Typical Operating Life 10 - 15 years in standard operating environment
Dimensions 115 x 130 x 71 mm (L, W, D) Mounting holes 105 x 75 mm
Enclosure Material ABS plastic
Cable Entry User Provided

Description

The GDA 4350 Carbon Dioxide Gas Sensor is a 3 wire Non Dispersive Infra red (NDIR) Carbon Dioxide (CO₂) sensor. It gives a 4-20 mA, 0-20 mA, 0-5 V or 0-10 V o/p over the range of the sensor.

Housed in an ABS plastic case with an aperture for the gas path. It uses laser LED to give reliable performance. Calibration for non critical safety applications is every 3 years. The low temperature version (suffix LT) operate from - 25°C to 30°C.


Highlighted Features

  • 24 VDC, 3 Wire Operation
  • 4-20 mA, 0-20 mA, 0-5 V or 0-10 V o/p linear over range of sensor
  • Ultra low cross sensitivity to other gas
  • Temperature and humidity compensated
  • Multiple ranges
  • No moving parts.
  • Factory calibrated
  • Low drift

Designed, manufactured and supported locally in Australia.