Holiday Closure: We will be closed from Mon, 23rd Dec & will return on Thurs, 2nd Jan 2025

Holiday Closure: We will be closed from Mon, 23rd Dec & will return on Thurs, 2nd Jan 2025

Search
Menu Close

Jumo tecLine Br - Bromine Sensor

Manufacturer:
The membrane-covered, amperometric sensor works on the three-electrode principle; it is used for the online measurement of hypobromous acid (HOBr). The electronic components integrated in the sensors provide a temperature-compensated current signal (4 to 20 mA) or a Modbus RTU output signal.
Part Number: 202637
Back to Previous Page
Call for pricing

Monitor Your Bromination Process Easily and Precisely with our Sensors

The membrane-covered, amperometric sensor works on the three-electrode principle; it is used for the online measurement of hypobromous acid (HOBr). The electronic components integrated in the sensors provide a temperature-compensated current signal (4 to 20 mA) or a Modbus RTU output signal.

The disinfectant bromine offers a greater disinfectant effect and higher pH value compared with chlorine compounds, while having a low corrosive effect. Abreacted bromine solutions irritate the mucous membranes to a lesser extent and have a neutral odor.

The membrane-covered, amperometric sensor works on the 3-electrode principle; it is used for online measurement hypobromous acid (HOBr). Suitable bromination methods include free bromine (HOBr) as well as 1-bromine-3-chlorine-5.5-dimethyl-hydantoin (BCDMH).

Depending on the type, the electronics integrated in the sensors provide a temperature-compensated current signal (4 to 20 mA) or a Modbus RTU output signal. A downstream device (indicator, controller, recorder, PLC, etc.) is used for calibration.

The sensor can be connected directly to indicating devices/transmitters/controllers. They supply the sensor with voltage and allow for easy calibration of the measuring system.

The sensor is not suitable for detecting the absence of bromine.


Features

  • 3-electrode principle
  • Easy calibration
  • Integrated temperature compensation
  • Proven measuring system
  • Electrical connection analog or digital
  • Connection to JUMO digiLine system optional


Application

  • Measurements in cooling water 
  • Swimming pool water
  • Service water
  • Process water
  • Seawater

Technical Information

Sensor Type

202637/61

(output signal 4 to 20 mA)

202637/71

(digital interface output signal)

Measurand Hypobromous acid (HOBr)

Area of application

Suitable bromination methods

Disinfection of drinking water, swimming pool water, service water, process water and seawater
free bromine (HOBr)
1-bromine-3-chlorine-5.5-dimethyl-hydantoin (BCDMH)
Measuring Principle Membrane-covered, amperometric, potentiostatic three-electrode system with integrated electronics
Measuring Cable Connection 2-pin terminal connection (2 × 1 mm2) 5-pin flange connector, M12
Voltage Supply UB DC 12 to 30 V
(galvanic isolation required)
UB DC 22.5 to 26 V
(galvanically isolated from the sensor)
Electromagnetic compatibilitya Interference emission: class Bb
Interference immunity: to industrial requirements
Output Signal 4 to 20 mA Modbus RTU
Burden/current Consumption ≤ (UB - 7.5 V) ÷ 0.02 A approx. 20 mA
Settling Time Approx. 2 h for initial startup
Inflow speed Approx. 15 cm/s
(corresponds to a flow of approx. 30 l/h when installed in the JUMO flow fitting (part no.: 00392611))
Measuring Rangesc 0.005 to 2 mg/l (ppm)
0.05 to 20 mg/l (ppm)
Resolution 0.001 mg/l with measuring range 2 mg/l
0.01 mg/l with measuring range 20 mg/l
Response time t90 approx. 2 minutes
Operating temperature
Measurement water 
ambience
0 to 45 °Cd
0 to 55 °C
Temperature compensation Automatic, using integrated temperature probee
pH value area of application pH 6.5 to pH 9.5
Zero point adjustment Not required
Slope adjustment On evaluation unit/controller using analytical determination
Disturbances Cl2: is also measured
ClO2: is also measured
O3: is also measured
Corrosion inhibitors and water hardness stabilizers can lead to measuring errors
Absence of the disinfectant max. 24 h
Pressure resistancef pabs max. 1.5 bar
prel max. 0.5 bar
Materials PVC-U, PEEK, stainless steel 1.4571
Dimensions Dia. 25 mm, length 220 mm Dia. 25 mm, length 205 mm
Weight Approx. 125 g

Maintenance

Inspection of the measuring signal

Replacement of membrane cap

Replacement of electrolyte

Regularly, at least once a week

Once a year (depending on the quality of the water)

Every 3 to 6 months

Storage

Sensor

Membrane Cap

Electrolyte

Can be stored indefinitely in a dry place, without electrolyte, between +5 and 40 °C

Can be stored indefinitely in the original packaging at +5 to 40 °C.

In original bottle, away from sunlight, for at least one year (or up to minimum best by date) at +5 to 35 °C.C

a EN 61326-1, EN 61326-2-3
b The product is suitable for industrial use as well as for households and small businesses.
c Other measuring ranges upon request.
d Prerequisite: no ice crystals in the measurement medium.
e Prerequisite: no temperature jumps in the measurement medium.
f No pressure fluctuations admissible. Pressure-free operation (atmospheric pressure) recommended.

Monitor Your Bromination Process Easily and Precisely with our Sensors

The membrane-covered, amperometric sensor works on the three-electrode principle; it is used for the online measurement of hypobromous acid (HOBr). The electronic components integrated in the sensors provide a temperature-compensated current signal (4 to 20 mA) or a Modbus RTU output signal.

The disinfectant bromine offers a greater disinfectant effect and higher pH value compared with chlorine compounds, while having a low corrosive effect. Abreacted bromine solutions irritate the mucous membranes to a lesser extent and have a neutral odor.

The membrane-covered, amperometric sensor works on the 3-electrode principle; it is used for online measurement hypobromous acid (HOBr). Suitable bromination methods include free bromine (HOBr) as well as 1-bromine-3-chlorine-5.5-dimethyl-hydantoin (BCDMH).

Depending on the type, the electronics integrated in the sensors provide a temperature-compensated current signal (4 to 20 mA) or a Modbus RTU output signal. A downstream device (indicator, controller, recorder, PLC, etc.) is used for calibration.

The sensor can be connected directly to indicating devices/transmitters/controllers. They supply the sensor with voltage and allow for easy calibration of the measuring system.

The sensor is not suitable for detecting the absence of bromine.


Features

  • 3-electrode principle
  • Easy calibration
  • Integrated temperature compensation
  • Proven measuring system
  • Electrical connection analog or digital
  • Connection to JUMO digiLine system optional


Application

  • Measurements in cooling water 
  • Swimming pool water
  • Service water
  • Process water
  • Seawater