Technical Information
|
Sensor Type
|
202637/61
(output signal 4 to 20 mA)
|
202637/71
(digital interface output signal)
|
Measurand |
Hypobromous acid (HOBr) |
Area of application
Suitable bromination methods
|
Disinfection of drinking water, swimming pool water, service water, process water and seawater free bromine (HOBr) 1-bromine-3-chlorine-5.5-dimethyl-hydantoin (BCDMH) |
Measuring Principle |
Membrane-covered, amperometric, potentiostatic three-electrode system with integrated electronics |
Measuring Cable Connection |
2-pin terminal connection (2 × 1 mm2) |
5-pin flange connector, M12 |
Voltage Supply |
UB DC 12 to 30 V (galvanic isolation required) |
UB DC 22.5 to 26 V (galvanically isolated from the sensor) |
Electromagnetic compatibilitya |
Interference emission: class Bb Interference immunity: to industrial requirements |
Output Signal |
4 to 20 mA |
Modbus RTU |
Burden/current Consumption |
≤ (UB - 7.5 V) ÷ 0.02 A |
approx. 20 mA |
Settling Time |
Approx. 2 h for initial startup |
Inflow speed |
Approx. 15 cm/s (corresponds to a flow of approx. 30 l/h when installed in the JUMO flow fitting (part no.: 00392611)) |
Measuring Rangesc |
0.005 to 2 mg/l (ppm) 0.05 to 20 mg/l (ppm) |
Resolution |
0.001 mg/l with measuring range 2 mg/l 0.01 mg/l with measuring range 20 mg/l |
Response time t90 |
approx. 2 minutes |
Operating temperature Measurement water ambience |
0 to 45 °Cd 0 to 55 °C |
Temperature compensation |
Automatic, using integrated temperature probee |
pH value area of application |
pH 6.5 to pH 9.5 |
Zero point adjustment |
Not required |
Slope adjustment |
On evaluation unit/controller using analytical determination |
Disturbances |
Cl2: is also measured ClO2: is also measured O3: is also measured Corrosion inhibitors and water hardness stabilizers can lead to measuring errors |
Absence of the disinfectant |
max. 24 h |
Pressure resistancef |
pabs max. 1.5 bar prel max. 0.5 bar |
Materials |
PVC-U, PEEK, stainless steel 1.4571 |
Dimensions |
Dia. 25 mm, length 220 mm |
Dia. 25 mm, length 205 mm |
Weight |
Approx. 125 g |
Maintenance
Inspection of the measuring signal
Replacement of membrane cap
Replacement of electrolyte
|
Regularly, at least once a week
Once a year (depending on the quality of the water)
Every 3 to 6 months
|
Storage
Sensor
Membrane Cap
Electrolyte
|
Can be stored indefinitely in a dry place, without electrolyte, between +5 and 40 °C
Can be stored indefinitely in the original packaging at +5 to 40 °C.
In original bottle, away from sunlight, for at least one year (or up to minimum best by date) at +5 to 35 °C.C
|